Show simple item record

dc.contributor.authorMeazea, AKM Moinul Haque
dc.date.accessioned2025-08-04T10:03:56Z
dc.date.available2025-08-04T10:03:56Z
dc.date.issued2022
dc.identifier.citationIraqi Journal of Scieneen_US
dc.identifier.urirepository.auw.edu.bd:8080//handle/123456789/910
dc.description.abstractModern radiotherapy facilities like 3-Dimensional conformal radio therapy (3DCRT), Intensity Modulated Radiotherapy (IMRT), were recently suggested in Co-60 machine with Multi-leaf Collimator (MLC). In this study, two reference chambers NE-2571#1205 and NE2581#537 were used for absolute dose measurement in Equinox accelerator. A comparison of dose measurement by two different IAEA protocols TRS-277 and TRS-398 has been studied. Analyzing TRS- 398, a common shaped empirical formula was developed for the four Co-60 units of four Medical Colleges in Bangladesh with fitting parameters. It was found that an average discrepancy in the determination of absorbed dose in water among the two different protocols TRS-277 and TRS-398 were 1.33 % for the chamber NE- 2571#1205 and 0.65 % for the chamber NE2581#537 with combined uncertainty ±1.59 % (k=1). A good convergence has been obtained in the concepts and methods in this study.en_US
dc.language.isoenen_US
dc.subjectCo-60 Teletherapy Units, Technical Report Series (TRS), Absorbed Absolute Dose.en_US
dc.titleMeasurement of Absorbed Dose in Water Using Co-60 gamma source Teletherapy Units According to IAEA Dosimetry Protocolsen_US
dc.typeArticleen_US