dc.contributor.authorMeaze, AKM Moinul Haque
dc.date.accessioned2025-07-31T07:53:30Z
dc.date.available2025-07-31T07:53:30Z
dc.date.issued2023
dc.identifier.urirepository.auw.edu.bd:8080//handle/123456789/904
dc.description.abstractAnalytic radiography is a normal image testing technique which has been utilized for quite a long time. It is recommended by specialists so they can identify any problem in patients' bodies without a cut. Thinking about its wide use, the principle objective of this investigation is to give a top notch picture by keeping the radiation portion as low as conceivable through identifying any variety in quality control (QC) boundaries. In this work, some standard quality control boundaries, for example, voltage exactness, time precision test, tube yield linearity, half value layer (HVL) of x-beam were measured. These quality control (QC) boundaries were estimated by a dosimeter keeping a distance of 100 cm from source. The voltage precision went from 0.31% to 4.67% and the time exactness test went from 0% to 2.29%. The consequences of this investigation show that all the QC boundaries are inside the acceptable level which guarantees the advancement of the low portion conveyed to the patients.en_US
dc.language.isoenen_US
dc.publisherSuan Sunandha Science and Technology Journalen_US
dc.subjectDiagnostic radiography, Quality control (QC), X-ray, HVL, Dosimeteren_US
dc.titleInvestigation of Some Regular X-ray Imaging Parameters in Suggestive Radiography of Four Hospitals in Bangladeshen_US
dc.typeArticleen_US