Investigation of Some Regular X-ray Imaging Parameters in Suggestive Radiography of Four Hospitals in Bangladesh
Abstract
Analytic radiography is a normal image testing technique which has been utilized for quite a long time. It is
recommended by specialists so they can identify any problem in patients' bodies without a cut. Thinking about its
wide use, the principle objective of this investigation is to give a top notch picture by keeping the radiation portion
as low as conceivable through identifying any variety in quality control (QC) boundaries. In this work, some
standard quality control boundaries, for example, voltage exactness, time precision test, tube yield linearity, half
value layer (HVL) of x-beam were measured. These quality control (QC) boundaries were estimated by a
dosimeter keeping a distance of 100 cm from source. The voltage precision went from 0.31% to 4.67% and the
time exactness test went from 0% to 2.29%. The consequences of this investigation show that all the QC
boundaries are inside the acceptable level which guarantees the advancement of the low portion conveyed to the
patients.
Collections
- 2023 [3]